Lead interaction: rare cause of oversensing during implantation procedure of implantable cardioverter-defibrillator system.

机译:引线交互作用:可植入式心脏复律除颤器系统在植入过程中过度感应的罕见原因。

获取原文
获取原文并翻译 | 示例
           

摘要

A variety of etiologies can cause erroneous detection in patients with implantable cardioverter defibrillator (ICDs). Interaction between two endocardial leads is rare and uncommon in causing electrical noise. During a reimplantation procedure of an ICD system in a 68-year-old man, additional electrical signals could be detected. The interaction between two endocardial defibrillation leads was identified as the cause of sensing problems. When it is not possible to extract the nonfunctional endocardial lead during implantation of the new electrode, it should be implanted away and not in parallel from the old one to avoid interaction between them. A variety of etiologies can cause erroneous detection in patiens with ICD. Interaction between two endocardial leads is rare and uncommon to cause electrical noise.
机译:多种病因会导致植入式心脏复律除颤器(ICD)患者错误检测。两种心内膜导联之间的相互作用很少,而且很少引起电噪声。在一名68岁的男性的ICD系统的再植入过程中,可以检测到其他电信号。两条心内膜除颤导线之间的相互作用被确定为感觉问题的原因。当在植入新电极期间无法取出无功能的心内膜导联时,应将其植入而不是与旧电极平行,以避免相互干扰。多种病因会导致ICD患者的错误检测。两根心内膜导线之间的相互作用很少见,并且很少引起电噪声。