Three different possible sources of mass fractionation have been investigated in an accelerator mass spectrometer system using sequential injection, namely: the influence of the accelerating/decelerating gaps in front of- and after the bouncing magnet, the influence of changes in the ion trajectories for the three carbon isotopes during bouncing and finally the influence of beam intensity. Measurements have indicated that the first two of these sources on the isotopic ratios measurements can be neglected. On the other hand, it was demonstrated that increasing beam intensity influences the ~(14)C/~(12)C ratio and to a lesser extent the ~(14)C/~(13)C ratio. This is probably related to space charge effects.
机译:在使用顺序注入的加速器质谱仪系统中,研究了三种不同的质量分级来源,即:弹跳磁体前后的加速/减速间隙的影响,离子弹道变化的影响。弹跳过程中的三个碳同位素,最后是束强度的影响。测量表明,同位素比测量中这些来源的前两个可以忽略。另一方面,已经证明增加的光束强度影响〜(14)C /〜(12)C比,并且在较小程度上影响〜(14)C /〜(13)C比。这可能与空间电荷效应有关。